Ion Chromatography System LTCGR8-1
Used in Food Testing, Chemical Industry, Beverage Testing, Drug testing, Forensic Science, Pharmaceutical, Molecular Biology, Medical, Research, Laboratory
- Built-in circulating 3D constant temperature technology:
- Temperature stability time is less than 30 mins, ensuring the accuracy and
- reliability of test data.
- The world's leading full-range series of ion chromatographic columns:
Price: $27,269.00
Download Catalog
Get Quote
Download Catalog
Get Quote
Features
- Built-in circulating 3D constant temperature technology:
- Temperature stability time is less than 30 mins, ensuring the accuracy and
- reliability of test data.
- The world's leading full-range series of ion chromatographic columns:
- High efficiency, large capacity of the columns for detecting ions of varied
- compositions.
- Self-Regenerating Electrolytic Micro-membrane Suppressor:
- High pressure resistance, small dead volume, highly responsive to signals.
- Auto-range Conductivity Detector:
- It can directly detect the signal from ppb to ppm without adjusting the range. Only one conductivity detector can detect anions and cations.
- Observatory intelligent workstation:
- Integrated control, intelligent start-up, shutdown and maintenance functions. Compatible with a variety of instruments.
Specifications
| Ion Chromatographic Pump: | |
| Maximum Pressure | 42 Mpa (Stainless steel) |
| Type | High-pressure and low-pulse two-piston tandem advection pump |
| Pressure Display Accuracy | ≤ 0.1 MPa |
| Flow Range | 0.001 ~ 9.999 mL/min |
| Pressure Fluctuation | ≤ 0.5% |
| Flow Stability | (0.2-0.5) mL/min ≤ 2%; (0.5-1.0) mL/min ≤ 1%; > 1.0 mL/min ≤ 1% |
| Manual Sample Injector: | |
| Contact Material of the Rotor | PEEK |
| Contact Material of Medium | PEEK/Ceramics |
| Column Heater: | |
| Operating Temperature Range | 20°C~60°C (68~140°F) |
| Controlling Temperature Accuracy | ± 0.01°C |
| Allowable Deviation of Column Heater’s Temperature | ± 1°C |
| Temperature Stability | ≤ 0.05°C/h |
| Conduction Detection System: | |
| Type | Constant temperature auto-range conductivity detector |
| Cell Volume | ≤0.8μL |
| Detection Range | 0~35000 μS/cm |
| Detection Resolution | ≤0.0020nS/cm |
| Output Voltage | -6000~+6000 mv (adjustable) |
| Electronic Noise | 0.02 nS |
| Baseline Noise | ≤ 0.001 μS/cm |
| Baseline Drift | ≤ 0.02μS |
| Operating Temperature Range | Room temperature +5°C~60°C(41~140°F) |
| Controlling Temperature Accuracy | ±0.01°C |
| Temperature Compensation | 1.7 %/°C |
| Maximum Pressure | 10.0 Mpa |
| Linear Range | ≥ 10 3 |
| Instrument Linearity | ≥0.999 |
| Quantitative Repeatability | ≤1.0% |
| Qualitative Repeatability | ≤0.5% |
| Minimum Detectable Concentration | Cl- ≤ 0.0005 ug/mL; Li+ ≤ 0.001 ug/mL; BrO3≤ 0.001 ug/mL |
| Flow System: | |
| Six-way Valve | PEEK material, pressure 5000 psi; Independent automatic collecting and flow function. |
| Suppressor: | |
| Type | Self-Regenerating electrolytic micro-membrane suppressor |
| Maximum Pressure | 6.0 Mpa |
| Dead Volume | <50 μL |
| Other Specifications: | |
| Dimension (LxWxH) | 350x470x510 mm |
| Net Weight | 26 kg |
| Gross Weight | 32 kg |
| Power | 150 W |